|
Click
here to download the entire case study (PDF)
"A Dynamic Site Investigation Adaptive Sampling and Analysis Program
for Operable Unit 1 at Hanscom Air Force Base"
Page 59
"To test the data generation software, comparisons were made between
the EnviroQuant and Ion Signature quantitative deconvolution software systems. Also shown in Tables 8-11
are the initial and continuing calibration line parameters and %RSD
RFcal and the percent difference as determined by the EnviroQuant software.
Note that all GC/MS total ion current chromatograms were acquired by
HP’s data acquisition system. Both Ion Signature and EnviroQuant, therefore, analyzed
the same data files. The tables reveal no statistical differences in
the information obtained. Moreover, the %RSD’s for the eighteen VOC
RFcal’s were well within the field criterion of 30% for two-thirds of
the target compounds and 40% for the remaining one-third of the organics
established for this field study. Good agreement was also found for
PCBs and PAH except for naphthalene. EnviroQuant produced a 65% RSD
RFcal in comparison to the 37% RSD calculated by Ion Signature. Excellent agreement
was also obtained between the two data analysis systems for the continuing
calibration data. The VOC and semivolatile data were typically within
the 30% required of fixed-based laboratory data."
"To determine instrument and method stability in the field, each
sample’s internal standard signal was compared against the corresponding
average internal standard signal calculated over the first two calibration
curves. The internal standard signal was inside the -50 to 150 percent
difference HAFB DQO for 134 of 157 samples (85% success rate) analyzed
by Ion Signature. EnviroQuant fell outside of the DQO for 32 of the same samples,
while missing the internal standard identity in two samples completely
for a 78% success rate. Both data analysis systems fell out of QC because
of coeluting matrix constituent(s) that affected the internal standard
integration."
Page 60
"Table 27 shows Ion Signature and EnviroQuant site sample data comparisons
for the same calibration files where at least one of the data analysis
systems identified a target compound above the reporting limit. Differences
between the two calculated concentrations were due to each data system’s
peak identification and integration routines. For example, Ion Signature provided
better integration where matrix constituents coeluted with either target
compounds or internal standards. EnviroQuant cut peaks in half depending
upon where peak separation was determined by the software. Target compound
concentrations in file #’s 100.D, 178.D, 179.D, and 194.D were reported
by Ion Signature whereas, EnviroQuant estimated their concentration below the
reporting limit. In addition, EnviroQuant missed compounds in file #'s
102.D, 143.D, and 178.D that were well above the reporting limit, with
Ion Signature missing one compound in file # 031.D. Of the 205 reported compounds
by both software systems, 85% had RPD’s 50% and 65% had RPD’s 20%. This
is remarkable given that both data systems identified peaks and integrated
signals very differently."
"Innovations
in Site Characterization"
Page xviii
"Cost Per
Sample:
Use of the Ion Signature
Software reduced GC/MS run times by a factor of 2-4, which increased
sample throughput, and thus decreased both direct costs (e.g., labor)
and indirect costs (by decreasing turn-around time of results within
the context of a dynamic workplan."
"Site-Specific
Accuracy/Precision Achieved:
The performance
of the Ion Signature software was compared with the data processing software typically
supplied with an HP GC/MS system. Of the 158 VOC samples analyzed, 4
samples required re-analysis by Ion Signature to resolve interference or linearity
issues, compared to 25 re-analyses required using the traditional software
package."
"The internal
standard signal was within the ± 50% difference acceptance range 85%
of the time for the Ion Signature software, compared to 78% for the other software."
"Software
Performance Information:
The Ion Signature software is capable of identifying low concentrations of target
analytes in the presence of high levels of matrix interferences. By
"seeing through" the matrix interferences, the software reduces
the need for sample re-analyses and dilution, and increases confidence
in surrogate, internal standard, and target compound identification
and quantification."
Page 12
"Most MS data analysis software packages can extract individual
ion fragments, however the underlying mathematical algorithms unique
to the Ion Signature software facilitate compound identification for these samples
regardless of matrix interferences. Ion Signature quantitative deconvolution software provides the technology necessary
to reduce long chromatographic run times from 30-40 minutes to 10 minutes."
Page 24
"Observations and lessons learned:
The Ion Signature quantitative deconvolution software can be used to obtain compound selectivity rather
than adjusting the gas chromatography operating conditions. This decreases
the per sample analysis time and increases the number of samples that
can be analyzed per day per instrument over standard GC/MS instruments.
The Software algorithms which "look through" non-target MS
ion signals can unambiguously determine compound identity, minimizing
masking of low-concentration target compounds by high-concentration
matrix interferents. In this context, low-level target compounds are
not lost because of the need to dilute the sample. Use of the Software
algorithms makes sample dilution less necessary, keeping detection limits
low and saving time and money."
|